The LI-830 and LI-850 are precision gas analyzers designed for continuous monitoring of CO₂ concentrations in ambient or process air. The LI-850 further integrates H₂O measurement for enhanced accuracy in humid environments. Both deliver dependable performance, user-friendly connectivity and versatile configurations — all with minimal maintenance.
Both models employ non-dispersive infrared (NDIR) technology within a temperature-controlled optical bench to provide reliable and stable readings over the full measurement range. The LI-830 measures CO₂ only; the LI-850 measures both CO₂ and water vapour in the same optical cell and compensates for H₂O effects on CO₂ readings for improved precision.
Available with optional internal pump and local display, they offer analog, current, serial and USB outputs for seamless integration in lab, environmental or OEM setups.
CO₂ measurement range: 0 to 20,000 ppm
LI-850 also measures H₂O: 0 to 60 mmol mol⁻¹
Output rate: up to 2 Hz
NDIR optical bench with low drift, high stability and noise <1 ppm
Multiple configurable analog and digital outputs (voltage, current, USB, RS-232)
Optional built-in pump for constant flow; optional onboard display for real-time read-out
Lightweight, low power consumption — suitable for remote or mobile deployments
Designed for long-term operation with minimal maintenance
Ambient air quality monitoring (indoor & outdoor)
Environmental and atmospheric research (CO₂ flux, sequestration studies)
Growth chambers, greenhouses and controlled-environment agriculture
Industrial process monitoring and OEM instrumentation integration
Drone/UAV-mounted CO₂ measurement systems
Leak detection, volcanology and carbon-pool monitoring
Technische specificaties
Technical Specifications
CO₂ Measurements (both models)
Range: 0 to 20,000 ppm
Accuracy: LI-850: within 1.5% of reading; LI-830: within 3% of reading
Zero drift: < 0.15 ppm/°C
Span drift: < 0.03%/°C
Total drift at 370 ppm: < 0.4 ppm/°C
RMS noise at 370 ppm (1-second filter): < 1 ppm
Lower detection limit: 1.5 ppm
Sensitivity to H₂O (LI-850 only): < 0.1 ppm CO₂ / mmol mol⁻¹ H₂O
H₂O Measurements (LI-850 only)
Range: 0 to 60 mmol mol⁻¹
Accuracy: better than 1.5% of reading
Drift at 0 mmol mol⁻¹: < 0.005 mmol mol⁻¹/°C
Span drift at 10 mmol mol⁻¹: < 0.006 mmol mol⁻¹/°C
Total drift at 10 mmol mol⁻¹: < 0.016 mmol mol⁻¹/°C
RMS noise at 10 mmol mol⁻¹ (1-second filter): < 0.01 mmol mol⁻¹
Lower detection limit: < 0.015 mmol mol⁻¹
General
Output rate: up to 2 measurements per second (2 Hz)
Response time (T90): CO₂: < 3.5 s (0 to 375 ppm); H₂O: < 3.5 s (0 to 21 mmol mol⁻¹)
Measurement principle: Non-Dispersive Infrared (NDIR)
Pressure compensation: 50 to 110 kPa
Maximum gas flow rate: 1 L/min
Power & Environment
Input voltage: 12–30 V DC
Power consumption: Without pump: ~4.0 W (0.33 A @12V); With pump: ~5.0 W (0.42 A @12V); during warm-up: up to ~14 W (1.2 A @12V)
Operating temperature: –20 °C to 45 °C
Relative humidity: 0 to 95% RH, non-condensing
Dimensions: 221 mm W × 146 mm D × 75 mm H
Weight: Without pump/display: ~1.0 kg; With pump/display: ~1.3 kg
Optional Features
Internal pump for consistent sample flow (nominal 0.75 L/min, 1 W consumption)
On-board monochrome display (400×200 px) showing CO₂, H₂O (LI-850), optical bench temperature and pressure
Configurable alarms for threshold triggers via analog or digital outputs