The Formaldemeter htV-M is a precise and versatile instrument for measuring formaldehyde in indoor air. With integrated data logging, temperature, and humidity measurement, it is ideal for reliable air quality analysis in laboratories, cleanrooms, and industrial environments.
Formaldehyde is a critical parameter in indoor air quality (IAQ), directly impacting health, safety, and product quality. The Formaldemeter htV-M provides a fast and reliable solution for detecting low concentrations of formaldehyde, even under varying temperature and humidity conditions.
Thanks to its electrochemical sensor and internal sampling pump, the instrument delivers direct readings, suitable for both spot checks and continuous monitoring.
The htV-M simultaneously measures:
Formaldehyde (ppm and mg/m³)
Temperature
Relative humidity
This provides a complete picture of air quality and allows better interpretation of measurement results in relation to environmental conditions.
The strength of the htV-M lies in its flexibility:
Handheld use for quick inspections
Continuous monitoring for up to 1 month with internal data logging
Real-time clock for time-stamped data
USB connectivity (optional WiFi) for data export
This makes the instrument suitable for both inspections and long-term validation and trend analysis.
✔ Instant readings at the push of a button
✔ Suitable for low-level formaldehyde detection
✔ Continuous logging and data analysis
✔ Compact and lightweight (±300 g)
✔ Reliable performance under varying environmental conditions
✔ Easy calibration
The Formaldemeter htV-M is widely used in:
Cleanrooms and controlled environments
Laboratories (food, pharma, R&D)
Healthcare (formaldehyde exposure monitoring)
Industrial production and quality control
Building management and IAQ assessments
Validation of HVAC and climate systems
Technical specifications
Technical specifications
Formaldehyde
Measurement range: 0 – 10 ppm (extendable)
Resolution: 0.01 ppm
Accuracy: ±10% at 2 ppm
Sensor type: electrochemical
Measurement
Sampling: internal pump (10 ml air sample)
Response time:
±60 sec (high accuracy mode)
±8 sec (fast mode)
Environmental parameters
Temperature: -40 to +128 °C
Relative humidity: 0 – 100% RH
General
Dimensions: 150 × 80 × 34 mm
Weight: ±300 g
Power supply: battery or mains
Interface: USB (optional WiFi)